EasyManua.ls Logo

Labitec CoaDATA 501 - 2.4.2 Error Messages

Labitec CoaDATA 501
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
CoaDATA 501 – Operators Manual 2.1 Page 36
PT documentation
Example: Conversion via Factor calibration curve in % and INR.
--- method store 1 ---
PT
actual date 02.12.99
cuv remove detection OFF
Automatic cuvette detection
mode DOUBLE
Double determination
coef.var = 5 %
Coefficient of variation
incubat = 60 s
Sample incubation time
start-reagent:
Lot 101xxx
Reagent lot no.
reagent = 100 ul
Reagent start volume
1st convers QUICK FACTOR
2nd convers INR
100% = 11.8s
factor = 52
ISI = 1.05
-----------------------
results:
PT
Method
patient _____________
Patient name
02.12.99, 12:58:38
Date, Time
time 1 = 11.7 s
1st measured time
time 2 = 12.1 s
2nd measured time
Mean = 11.9 s
Mean of measured times
INR = 1.01
Conversion to INR
quick = 98.1%
Conversion to PT %