ALARM? Input Alarm Parameter Query
Input: ALARM?[term]
Returned: <off/on>, <high value>, <low value>, <deadband>, <latch
enable> [term]
Format: n,+nnn.n,+nnn.n,+nn.n,n (Refer to command for description)
ALMRST Reset Alarm Status Command
Input: ALMRST[term]
Remarks: Clears both the high and low status of the alarm, including
latching alarm.
ANALOG Analog Output Parameter Command
Input: ANALOG <mode>, <range>[term]
Format: n,n
<mode> Specifies mode in which analog output operates
where 0 = voltage mode and 1 = current mode.
<range> Sets temperature range that analog output uses as
full scale.
0 = 0 – 20 K 3 = 0 – 325 K
1 = 0 – 100 K 4 = 0 – 475 K
2 = 0 – 200 K 5 = 0 – 1000 K
Example: ANALOG 0,1[term] – Sets analog output to voltage mode
(0 – 10V) 100.0 K at +100% output (+10.0 V) and 0.0 K at
0% output (0.0 V).
ANALOG? Analog Output Parameter Query
Input: ANALOG?[term]
Returned: <mode>, <range> [term]
Format: n,n (Refer to command for definition)
AOUT? Analog Output Data Query
Input: AOUT?[term]
Returned: <analog output>[term]
Format: +nnn.nn
Remarks: Returns the percentage of output of the analog output.