Lake Shore Model 211 User’s Manual
4-
16
Remote Operation
BRIGT Display Brightness Command
Input: BRIGT <bright>[term]
Format: nn
<bright> Sets display brightness. Valid entries: 0 15,
0 = least bright, 15 = most bright. Default = 8.
BRIGT? Display Brightness Query
Input: BRIGT?[term]
Returned: <bright>[term]
Format: nn (Refer to command for description)
CRDG? Celsius Reading Query
Input: CRDG?[term]
Returned: <temp value>[term]
Format: ±nnnnnn
Remarks: Also see the RDGST? command.
CRVDEL Curve Delete Command
Input: CRVDEL <curve>[term]
Format: nn
<curve> Specifies user curve to delete. Only valid entry is
21. (Curve number is used to retain compatibility with
existing instrument line. Curve number 21 must be sent
with the command or else the command will be ignored.)
CRVHDR Curve Header Command
Input: CRVHDR <curve>, <name>, <SN>, <format>,
<limit value>, <coefficient>[term]
Format: nn,aaaaaaaaaaaaaaa,aaaaaaaaaa,n,+nnn.nnn,n
<curve> Specifies user curve. Valid entry: 21.
<name> Curve name. Limited to 15 characters.
<SN> Curve serial number. Limited to 10 characters.
<format> Curve data format. Valid entries:
2 = V/K, 3 = /K, 4 = log /K.
<limit value>Curve temperature limit in kelvin (Unused).
<coefficient> Curves temperature coefficient.
Valid entries: 1 = negative, 2 = positive.
Remarks: Configures the user curve header.