Lake Shore Model 211 User’s Manual
Remote Operation 4-
19
INCRV Input Curve Number Command
Input: INCRV <curve number>[term]
Format: nn
<curve number> Specifies which curve the input uses. If
specified curve parameters do not match
the input, the curve number defaults to 0.
Valid entries: 0 = none, 1 20 = standard
curves, 21 = user curve.
Remarks: Specifies curve the input uses for temperature conversion.
Example: INCRV 21[term] The input User Curve 21 for
temperature conversion.
INCRV? Input Curve Number Query
Input: INCRV?[term]
Returned: <curve number>[term]
Format: nn (Refer to command for description)
INTYPE Input Type Parameter Command
Input: INTYPE <sensor type>[term]
Format: n
<sensor type> Specifies input sensor type. Valid entries:
0 = Silicon Diode 3 = 100 Platinum/500
1 = GaAlAs Diode 4 = 1000 Platinum
2 = 100 Platinum/250 5 = NTC RTD
Example: INTYPE 0[term] Sets input sensor type to silicon diode.
INTYPE? Input Type Parameter Query
Input: INTYPE?[term]
Returned: <sensor type>[term]
Format: n (Refer to command for description)
KEYST? Keypad Status Query
Input: KEYST?[term]
Returned: <keypad status>[term]
Format: n 1 = key pressed, 0 = no key pressed.
Remarks: Returns keypad status since the last KEYST?. KEYST?
returns 1 after initial power-up.