6300 Rev H User’s Manual
125-9106A Page 79
Command Description
Data sent from
PC
Data sent from
6300
Pm Read the value for Pitot offset m, where m is as
defined in the table below.
m Pitot offset
1 10% offset
2 20% offset
3 30% offset
4 40% offset
5 50% offset
6 60% offset
7 70% offset
8 80% offset
9 90% offset
None Pitot offset value
Pm=n Set the Pitot offset m to a value of n. This
command requires the calibration password to be
set.
n 0 if OK; or
Error code
SE Returns details of 13 Self-test parameters in the
following format: parameter code=n, where n=0
(pass) or 1 (fail). The parameter codes are
explained in the table below.
Param.
code
Description
A Ps control solenoid
B Ps output solenoid
C Pt control solenoid
D Pt output solenoid
E Pressure pump
F Vacuum pump
G Ps control system
H Pt control system
I Ps main Xd measurement
J Pt main Xd measurement
K Ps control Xd measurement
L Pt control Xd measurement
M Baro Xd measurement
None 13 parameters
with pass/fail
code for each