6300 Rev H User’s Manual
125-9106A Page 78
Command Description
Data sent from
PC
Data sent from
6300
SO=n Returns the values for all the Static (n=0) or Pitot
(n=1) non-linearity offsets as shown below.
Static offsets Pitot offsets
S0% =x.xxxx
S5% =x.xxxx
S10%=x.xxxx
S15%= x.xxxx
S20%= x.xxxx
S25%= x.xxxx
S30%= x.xxxx
S35%= x.xxxx
S40%= x.xxxx
S45%= x.xxxx
S50%= x.xxxx
S55%= x.xxxx
S60%= x.xxxx
S65%= x.xxxx
S70%= x.xxxx
S75%= x.xxxx
S80%= x.xxxx
S85%= x.xxxx
S90%= x.xxxx
S95%= x.xxxx
S100%= x.xxxx
P0% =x.xxxx
P5% =x.xxxx
P10%=x.xxxx
P15%= x.xxxx
P20%= x.xxxx
P25%= x.xxxx
P30%= x.xxxx
P35%= x.xxxx
P40%= x.xxxx
P45%= x.xxxx
P50%= x.xxxx
P55%= x.xxxx
P60%= x.xxxx
P65%= x.xxxx
P70%= x.xxxx
P75%= x.xxxx
P80%= x.xxxx
P85%= x.xxxx
P90%= x.xxxx
P95%= x.xxxx
P100%= x.xxxx
n Static or Pitot
offset values
Sm Read the value for Static offset m, where m is as
defined in the table below.
m Static offset
1 10% offset
2 20% offset
3 30% offset
4 40% offset
5 50% offset
6 60% offset
7 70% offset
8 80% offset
9 90% offset
None Static offset
value
Sm=n Set the Static offset m to a value of n. This
command requires the calibration password to be
set.
n 0 if OK; or
Error code