6300 Rev H User’s Manual
125-9106A Page 77
Command Description
Data sent from
PC
Data sent from
6300
Cm Read the value for calibration point m, where m is
defined in the table below.
m Calibration point
1 Pitot vacuum point
2 Pitot full-scale point
3 Static vacuum point
4 Static full-scale point
None Value for
calibration point
m
Cm=n Set the calibration point m to the value n. This
command required the calibration password to be
set.
n 0 if OK; or
Error code
CC=n Calibration complete (n=1)
Calibration Void (n=0)
After this command is sent, the calibration
password is reset. The calibration password is
also reset on power-up
n 0 if OK; or
Error code
PZ Read the zero-offset value for the Pitot transducer None Pitot zero-offset
value
PZ=n This requires the calibration password to be set.
Set the zero-offset value for the Pitot transducer to
n
n 0 if OK; or
Error code
PS Read the slope value for the Pitot transducer None Pitot slope value
PS=n This requires the calibration password to be set.
Set the slope value for the Pitot transducer to n
n 0 if OK; or
Error code
SZ Read the zero-offset value for the Static
transducer
None Static zero-offset
value
SZ=n This requires the calibration password to be set.
Set the zero-offset value for the Static transducer
to n
n 0 if OK; or
Error code
SS Read the slope value for the Static transducer None Static slope
value
SS=n This requires the calibration password to be set.
Set the slope value for the Static transducer to n
n 0 if OK; or
Error code
HO Read the value of ‘Hours of Operation None Value of Hours
of operation
LC Read the ‘Last calibration date’. The string
returned is in the format ‘mm/dd/yy’
None Date as 8-byte
string
LC=n This command requires the calibration password
to be set. Set the ‘Last calibration date’ to n,
where n has the format ‘mmddyy’ (no ‘/’ in
between)
n 0 if OK; or
Error code
NC Read the ‘Next calibration due’ date. The string
returned shows ‘mm/dd/yy’.
None Date as 8-byte
string
OC=n Clears all offset values to zero for Static offsets
(n=0) or Pitot (n=1). This command requires the
calibration password to be set.
n 0 if OK; or
Error code