Lighthouse ApexPortable Operating Manual
A-6 248083440-1 Rev 5
Additional status bits are shown in 40056 displayed in Table A-5.
12 ApexPortable:
FLOW STATUS: Set to 1 when unit’s flow is out of
spec, else set to 0.
13 ApexPortable:
SERVICE STATUS: Set to 1 when unit needs to be
serviced, else set to 0.
14 ApexPortable:
THRESHOLD HIGH STATUS: Set to 1 when unit’s
high alarm threshold is exceeded, else set to 0.
15 ApexPortable:
THRESHOLD LOW STATUS: Set to 1 when unit’s
low alarm threshold is not met, else set to 0.
Table A-5 Additional Device Status (40056)
Bit Description
0 ApexPortable:
LASER POWER STATUS: Set to 1 when unit’s
LASER current is out of spec, else set to 0
1 ApexPortable:
LASER CURRENT STATUS: Set to 1 when unit’s
LASER power is out of spec, else set to 0
2 ApexPortable:
LASER SUPPLY STATUS: Set to 1 when unit’s
LASER supply is out of spec, else set to 0
3 ApexPortable:
LASER LIFE STATUS: Set to 1 when unit’s LASER
supply is out of spec, else set to 0
4 ApexPortable:
NO FLOW STATUS: Set to 1 when unit’s flow is below
no flow threshold causing LASER to turn off, else set to 0
5 ApexPortable:
PHOTOAMP SUPPLY STATUS: Set to 1 when unit’s
photoamp supply is out of spec, else set to 0
Table A-4 Current Device Status (40003/40057)
Bit Description