248083440-1 Rev 5 2-3
ApexP3
Specifications
Size Range 0.3 - 5.0 μm
Channel Thresholds Standard 2-channel: 0.3, 0.5μm; 4-channel:
0.3, 0.5, 1.0, 5.0μm
Flow Rate 1.0 CFM (28.3 LPM)
Counting Efficiency Meets ISO 21501-4
Data Storage Rotating Buffer, 3000 records
Light Source LASER diode
External Connections RS485 RJ45 (RS485 and RS232 data); DIN
(factory); Ethernet RJ45 (Ethernet data);
calibration port
Zero Count Level <1 count/5 minutes (per ISO 21501-4)
Max ISO Probe tubing Maximum tubing length to ISO = 16 feet
Calibration NIST Traceable
Communication Modes MODBUS ASCII; MODBUS RTU;
MODBUS TCP
Supporting Software LMS Pharma/Pro v7.3.1 or higher, LMS
Express 7.5 or higher
Operating Altitude Sea Level to 2000 meters (6561.66 feet)
Power Input
Requirements
24VDC ±5% @ 5.0A, 120W
Power Supply
Specifications
100-240VAC, 47-63Hz 1.4A input, 24VDC
5.0A 120W max output
Enclosure Stainless Steel, VHP compatible
Dimensions
9.00” (w) x 8.46” (h) x 6.21” (d) [22.86 x 21.49 x
15.77 cm]
Weight
10.0 lbs (4.54 kg)
Operating Temp/RH 50° F to 104° F (10° C to 40° C) / 20% to 95%
non-condensing at maximum 2000M altitude
Storage Temp/RH 14° F to 122° F (-10° C to 50° C) / Up to 98%
non-condensing
Table 2-1 ApexP3 Specifications