Lighthouse ApexPortable Operating Manual
5-52 248083440-1 Rev 5
Data and settings cannot be changed via the instrument web page. The
ApexPortable must be set up using the Ethernet interface and correct
IP for the network in which it is installed. Serial-configured (RS485
LAN) ApexPortable instruments do not support the web page feature.
The web page feature requires a valid TCP/IP address to be assigned to
the
ApexPortable.
Generate a Report From Data
Figure 5-84 Reports Main Screen
From the MAIN screen, press REPORTS.
1. Select the Report desired: FED STD FT3, ISO 14644-1 or EU
GMP 2009.
2. Press REPORT ENABLE to turn on green check (Figure 5-84).
3. Press the REPORT button, which will display the saved reports
list.
Figure 5-85 Reports Screen
4. Press ADD, which displays the Data View Buffer screen.