Operate the ApexPortable
248083440-1 Rev 5 5-49
• When "REPORT" is enabled, the "LOC" and "SAVE" buttons are
displayed only when there are at least the minimum number of
locations needed for the recipe area available.
• If REPORT is not active or the number of locations do not meet the
minimum needed in setup, the LOC and SAVE buttons are not
displayed. See Figure 5-79.
Figure 5-79 Report Setup Screen - No Locations
The next section contains an example of setting up the instrument to
run a report.
Set Up Counter to Run Reports
For this example, the FED STD FT
3
report will be used. See Figure 5-
80.
Figure 5-80 FED STD FT3 Report Screen
• Set the instrument for the Fed Std ft
3
report, 100 ft
2
, Class 100,
unidirectional air flow, In Operation room status.
• This will generate the following minimum requirements:
• Min Locations: 2
• Min Samples: 5