Lighthouse ApexPortable Operating Manual
5-50 248083440-1 Rev 5
Note:
The Minimum
Volumes are measured in the
standard measurement
required by each specific
report.
The number of digits allowed
for area to calculate volume is
limited to a 5 digit value for ft2
and a 4 digit value for m2.
Based on the minimum requirements displayed in Table 5-1, configure
the instrument as follows:
Press MAIN to return to the MAIN screen.
Press CONFIG to enter the Configuration screen.
Press SAMPLE.
Note:
The Sample time
automatically changes to
31:42 (31 minutes, 42
seconds).
Press the CYCLES button.
Set the CYCLES value for 1 and press ENTER. Note that the
instrument beeps twice and clears the value. This is intentional as
the minimum cycles is 2 and changing to 1 will cause errors.
Entering more than 2 will work, though, if needed.
Once the primary settings of Class and Area are entered, the values
are set for sample time/volume and cycles. Unless required by
SOP, there is no need to change these other values.
Make sure to add the needed locations and press BACK.
Note:
This is all that is
needed to meet the minimum
requirements to classify the
report.
Increase any of the
configuration parameters in
order to meet additional
reporting needs.
Press the MAIN button.
Press START and allow the instrument to record the required
sample(s) at the first location.
Figure 5-81 Prompt to Change LOCation
When it has finished, it will prompt to change to next location; a
YES response will change the Location but the user must press the
START button when set up and ready in the next Location.
Move the instrument to the second location and press START.
Table 5-1 Minimum Volume per Channel Size
Ch Size (μ)Min Vol (ft
3
)
0.3 0.100
0.5 0.200
1.0 0.919
5.0 31.698