Lighthouse ApexPortable Operating Manual
5-38 248083440-1 Rev 5
Diff, Diff / ft3, Diff / m3 - Indicate that each channel size’s data is
displayed in differential mode. If a channel was disabled, then there
is a blank space in its column; # indicates raw counts; ft
3
and m
3
indicate normalized counts.
Cuml, Cuml/ft3, Cuml/m3 - Indicates that each channel size’s
data is displayed in cumulative mode. If a channel was disabled,
there will be a blank space in this column. # indicates raw counts
while ft
3
and m
3
indicate normalized counts.
Loc - Indicates the location at which the data record was recorded.
The Location name listed is the alphanumeric label that was saved.
Smpl - Indicates the Sample Time (HH:MM:SS) at which the data
record was sampled.
Date - Indicates the instrument date on which the data was
recorded. The date will display in the format selected in the
configuration (MM/DD/YYYY, DD/MM/YYYY or YYYY/MM/
DD).
Time - Indicates the instrument time at which the data was
recorded (HH:MM:SS).
Instr - Indicates the state of the instrument at the time the data was
recorded. Instrument states include GOOD or SRVC.
Flow - Indicates the flow state of the instrument at the time the data
record was recorded. Flow is recorded as OK or ALRM. If the flow
was in alarm, it will print as Flow: Alert on the printouts.
Alarm - Indicates (NONE or YES) if the data record exceeded any
of the alarm thresholds of any of the channels that were enabled for
alarms. If the data record exceeded alarm thresholds, printouts
show "Alarm: Yes".
Laser - Indicates the status of the laser at the time the data record
was recorded; possible laser states include OK or SRVC. If the laser
needs to be serviced, it will print as Laser: Service.
If SRVC appears, printouts will say “Service Required”; the sensor
may need cleaning or repair. Please contact Lighthouse Technical
Support at 1-800-945-5905 (USA Toll Free) or 1-541-770-5905
(Outside of USA).