20/37
To charge the battery:
Connect the SCU to the mains.
Turn on the SCU.
While the battery is being charged, the status indicator will indicate the battery charge status (see table below).
In case of a power failure or if the SCU is disconnected from the mains without being turned off, the status indica- tor will
continue indicating the battery charge status.
Audio alarm signals will indicate any status change. After sounding 5 times, the audio alarm signals will stop.
To mute the audio alarm signal:
Mute audio alarm signal (see „Mute“).
Power Indicator Battery Status Indicator Audio Alarm Signal Power Supply
Off Off No signal No indication
Off Off Power fail signal Power supply insufcient
No mains supply
No battery or battery charge level below
25%
Connect SCU to mains.
-or-
Charge battery.
On (Green) Off No signal Mains supply ok
No battery or battery fully charged
On (Green) Flashing quickly (yellow) No signal No mains supply
Battery in use
Battery fully charged
On (Green) Flashing quickly (yellow) No signal Mains supply ok
Battery not in use
Battery charge level between 50% and 90%
On (Green) Flashing quickly (yellow) Every 15 minutes No mains supply
Battery in use
Battery charge level between 50% and 90%
On (Green) Flashing quickly (yellow) No signal Mains supply ok
Battery not in use
Battery charge level between 25% and 50%
On (Green) Flashing quickly (yellow) Every 5 minutes No mains supply
Battery in use
Battery charge level between 25% and 50%
On (Green) Flashing quickly (yellow) No signal Mains supply ok
Battery not in use
Battery charge level below 25%
On (Green) Flashing quickly (yellow) Every minute No mains supply
Battery in use
Battery charge level below 25%