10.10 CPR/Air Disconnected Mode
SCU activates CPR/Air Disconnected mode automatically if air pipe is not connected correctly or removed while in use.
When CPR/Air Disconnected mode is activated:
Indicator 8 lights red.
Uninterrupted audio alarm signal sounds.
Mattress deates.
To stop audio alarm signal:
Connect air pipe correctly.
-or-
Switch off SCU using green illuminated power switch on side of SCU.
Mute audio alarm signal (see Mute).
10.11 System error
If a system error occurs:
Indicator 9 ashes red.
Audio alarm signal (synchronized with ash sequence) sounds.
NOTE Thenumberoftimesthatthealarmsoundsandindicator9ashesindicatesthetypeoferror.Thisinformation
will be required when contacting a service engineer.
To mute audio alarm signal:
Mute audio alarm signal (see Mute).
To stop audio alarm signal:
Reset system by switching SCU off and back on.
NOTE The error signal will repeat every time the device is switched on until the cause of the error is reme- died.
10.12 Service
LINET® recommends maintenance after 10,000 operating hours of the SCU. Indicator 12 lights yellow after 10,000 ope-
rating hours.
When indicator 11 lights yellow:
Notify LINET® Service.
WARNING!
Risk of injury due to incorrect positioning of patient!
Ü
If a system error occurs, do not continue to use mattress replacement system and notify LINET® Service imme-
diately.