11.3 Transport Mode
Transport mode allows disconnecting the mattress from the SCU for transporting the bed without air escaping from the
mattress for up to 24 hours.
NOTE In Transport mode, alternating pressure therapy is not possible.
To activate Transport mode:
Select MAX mode in order to completely inate all mattress cells (see MAX Mode).
Disconnect air connector plug and rotate its end and body in opposite directions until red symbol ▲ and
black symbol ● are aligned and no air is escaping.
11.4 Power Failure
In case of a power failure, the mattress will remain inated for at least 24 hours. However, alternating pressure therapy is
not possible.
If no power supply will be available for more than 24 hours:
Switch to Transport mode (see Transport Mode).
WARNING!
Risk of injury due to incorrect positioning of patient!
Ü
Prior to more than 30 minutes of Transport mode, consult qualied hospital staff.
WARNING!
Risk of injury due to power failure!
Alternating pressure therapy is not possible during power failures.
Ü
In case of power failure, seek medical advice immediately.
Fig. 15 Transport Mode