6.3 Conductivity measurement
The conductivity measurement is a comparatively uncomplicated measurement. When used
properly the standard electrodes are stable for a long time and can be calibrated with the
integrated Cal function.
Attention: The device covers a very wide range of measurements. However a suitable elec-
trode for the measurement range must be used.
Range
CELL
- rAnG
1 2 3 4 5
0.01 0.000 -
5.000 µS/cm
0.00 - 50.00
µS/cm
0.0 - 500.0
µS/cm
0 - 5000
µS/cm
0.00 - 50.00
mS/cm
0.1 0.00 - 50.00
µS/cm
0.0 - 500.0
µS/cm
0 - 5000
µS/cm
0.00 - 50.00
mS/cm
0.0 - 500.0
mS/cm
1 (see scope
of delivery)
0.0 - 500.0
µS/cm
0 - 5000
µS/cm
0.00 - 50.00
mS/cm
0.0 - 500.0
mS/cm
0 - 1000
mS/cm
10 0 - 5000
µS/cm
0.00 - 50.00
mS/cm
0.0 - 500.0
mS/cm
0 - 1000
mS/cm
---
If the range selection is set to "Auto Range", the range with the best resolution is automati-
cally selected. However the logger or interface operation requires a fixed pre-selection of the
measurement range from the above table (no logger/interface operation with Auto Range!).