6.3 Conductivity measurement
The conductivity measurement is a comparatively uncomplicated measurement. When used
properly the standard electrodes are stable for a long time and can be calibrated with the
integrated Cal function.
Attention: The device covers a very wide range of measurements. However a suitable elec-
trode for the measurement range must be used.
Range
CELL
- rAnG
1 2 3 4 5
0.01 0.000 -
5.000 µS/cm
0.00 - 50.00
µS/cm
0.0 - 500.0
µS/cm
0 - 5000
µS/cm
0.00 - 50.00
mS/cm
0.1 0.00 - 50.00
µS/cm
0.0 - 500.0
µS/cm
0 - 5000
µS/cm
0.00 - 50.00
mS/cm
0.0 - 500.0
mS/cm
1 (see scope
of delivery)
0.0 - 500.0
µS/cm
0 - 5000
µS/cm
0.00 - 50.00
mS/cm
0.0 - 500.0
mS/cm
0 - 1000
mS/cm
10 0 - 5000
µS/cm
0.00 - 50.00
mS/cm
0.0 - 500.0
mS/cm
0 - 1000
mS/cm
---
If the range selection is set to "Auto Range", the range with the best resolution is automati-
cally selected. However the logger or interface operation requires a fixed pre-selection of the
measurement range from the above table (no logger/interface operation with Auto Range!).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lovibond SD 320 Con and is the answer not in the manual?

Lovibond SD 320 Con Specifications

General IconGeneral
BrandLovibond
ModelSD 320 Con
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals