16. Error and system messages
Error messages for the measurement
Meaning Recommendation
No display or confused
symbols
Device does not react to
pushing buttons
Battery is depleted Insert new battery
System error Disconnect battery and
mains adapter, wait a short
time and plug in again
Device defective Send in for repair
Err.1
Below measurement range Check: Is the measurement
above the permissible
measurement range of the
sensor? -> Measurement
value is too high!
Sensor defect Send in for repair
Err.2
Measurement range
undercut
Check: Is the measurement
below the permissible
measurement range of the
sensor? -> Measurement
value is too low!
Sensor defect Send in for repair
Err.3
at alarm threshold 1
Display range exceeded Pressing
of
or
Err.7
System error Send in for repair
Results significantly outside
detection range
Check: Is the measurement
within the permissible
measurement range of the
sensor?
---
Display value not calculable
Measurement range or input
parameter exceeded
Check measurement range
Measurements too unstable Wait for reading to stabilize
> CAL <
CAL blinks in the upper
display
The pre-adjusted calibration
interval has lapsed or the
last calibration was invalid
The device must be
calibrated or inactivation of
c.INT = OFF
no Auto
LOGG rAnG
Logger could not be started Autorange for the display
range is activated
=> Setting in the
configuration menu
Error messages of the automatic calibration
CAL Err.1 Cell constant too high Determined constant may
not exceed 1.2
.
cell range
CAL Err.2 Cell constant too low Determined constant may
not be below 0.4
.
cell range
CAL Err.3 Solution in the incorrect
range
Incorrect cell range /
incorrect solution / far
outside of tolerance
CAL Err.4 Temperature incorrect Outside of permissible
temperature range: 0.0 –
34.0 °C (or 0.0 – 27.0 °C at
111.8 mS/cm)
If "bAt" blinks in the display, the battery is very low. Measurement can continue for a brief
time. If "bAt" remains in the display, the battery is fully depleted and must be replaced. No
more measurements can take place.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lovibond SD 320 Con and is the answer not in the manual?

Lovibond SD 320 Con Specifications

General IconGeneral
BrandLovibond
ModelSD 320 Con
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals