3-2 Rad-57 Signal Extraction Pulse CO-Oximeter Operator’s Manual
3
Monitor Setup
INITIAL SETUP
1. Inspect the oximeter case for damage.
2. Install 4 (four) new AA alkaline batteries.
3. Turn the device on, the LEDs will scroll in the display window as the sensor
calibrates, verify all indicators illuminate and speaker sounds a brief tone.
4. Configure the device for your regional power line frequency (50 or 60 hz) if needed.
Default is 60 hz (standard for the United States). See Section 4, Special Menu,
Special menu - Line Frequency Configuration.
CAUTION: THE DEVICE MUST BE CONFIGURED TO MATCH YOUR LOCAL POWER
LINE FREQUENCY TO ALLOW FOR THE CANCELLATION OF NOISE
INTRODUCED BY FLUORESCENT LIGHTS AND OTHER SOURCES.
No other setup is required. Refer to Section 4, General Setup and Use for additional steps
to verify proper functioning of the device.