Rad-57 Signal Extraction Pulse CO-Oximeter Operator’s Manual 4-3
4
operation
DEFAULT SETTINGS
The Rad-57 is shipped configured with all values set to factory default. Adjustments made
by the user will be retained after a power cycle for all values except alarm silence.
NOTE: Before use, confirm that the device settings are set appropriately.
The following table outlines the default values and the values the Rad-57 reverts to after
a power cycle:
OPTION FACTORY DEFAULT SETTING
CONFIGURABLE SETTING/
BEHAVIOR AT POWER-UP
SpO
2
high alarm limit (Default is OFF)
Set to Pre-Power Down Setting
See Section 5, Alarm Limits for
all settings.
SpO
2
low alarm limit (Default is 90%)
Pulse rate high alarm limit (Default is 140 BPM)
Set to Pre-Power Down Setting
See Section 5, Alarm Limits for
all settings.
Pulse rate low alarm limit (Default is 50 BPM)
SpCO* high alarm limit (Default is 10%)
Set to Pre-Power Down Setting
See Section 5, Alarm Limits for
all settings.
SpCO* low alarm limit (Default is OFF)
SpMet* high alarm limit (Default is 3%)
Set to Pre-Power Down Setting
See Section 5, Alarm Limits for
all settings.
SpMet* low alarm limit (Default is OFF)
Averaging Time (Default is 8 seconds)
Set to Pre-Power Down Setting
2, 4, 8, 10, 12, 14, or 16 seconds
FastSat (Default is OFF)
Set to Pre-Power Down Setting
Off/On
Sensitivity (Default is Normal)
Set to Pre-Power Down Setting
Normal, MAX or APOD
Display brightness (Default is level 2)
Set to Pre-Power Down Setting
Levels 1 through 4
Pulse tone volume (Default is level 4)
Set to Pre-Power Down Setting
Levels 1 through 7
Alarm Volume (Default is level 1)
Set to Pre-Power Down Setting
Levels 1 through 4
Line Frequency (Default is 60 Hz)
Set to Pre-Power Down Setting
50 Hz/ 60 Hz
Trend Active (Default is ON)
Set to Pre-Power Down Setting
Off/On
Alarm Silence (Default is All Alarms Active)
Set to All Alarms Active
Alarm on/off
*For each corresponding parameter, refer to the applicable device.