Rad-57 Signal Extraction Pulse CO-Oximeter Operator’s Manual 7-1
7
specifications
Rad-57 Family Specifications
PERFORMANCE
Measurement range
Oxygen Saturation (%SpO
2
): 1-100%
Carboxyhemoglobin Saturation (%SpCO*) 0 - 99%
Methemoglobin Saturation (%SpMet*) 0 - 99.9%
Pulse Rate (bpm): 25-240 beats per minute
Perfusion Index: 0.02 - 20%
ACCURACY
Saturation
70% to 100%
No Motion
1
Adults, Pediatrics ±2 digits
Neonate ±3 digits
Motion
2
Adults, Pediatrics ±3 digits
Neonate ±3 digits
Low Perfusion
3
Adults, Pediatrics ±2 digits
Neonate ±3 digits
Carboxyhemoglobin Saturation Accuracy (%SpCO*)
7
1% - 40% ±3 digits
Methemoglobin Saturation Accuracy (%SpMet*)
8
1% - 15% ±1 digit
Pulse Rate (bpm) during no motion conditions
1
Adults, Pediatrics, Neonate 25 to 240 ±3 digits
Pulse Rate (bpm) during motion conditions
2
Adults, Pediatrics, Neonate 25 to 240 ±5 digits
Resolution
Oxygen Saturation (%SpO
2
)1%
Carboxyhemoglobin saturation (%SpCO*), digital display 1%
Carboxyhemoglobin saturation(%SpCO*), continuous bar display
3, 6, 9, 12, 15, 20, 25,
30, 35, > 40%
Methemoglobin saturation (%SpMet*), digital display .1%
Methemoglobin saturation(%SpMet*), continuous bar display .5, 1-5, 7.5, 10, 15 >20%
Pulse Rate (bpm)
9
1 bpm
*See Model Summary for applicable device.