4-14 Rad-57 Signal Extraction Pulse CO-Oximeter Operator’s Manual
4
operation
Trend Setup and Use
INTRODUCTION
The Rad-57 can store 72 hours of SpO
2,
Pulse Rate, SpCO*, SpMet* and Perfusion Index
trend data, captured at 2 second intervals. This trend data can then be transferred to a
PC for evaluation.
Trend data is stored in non-volatile memory, so it is not erased when the device is shut off
or when the batteries are replaced.
A Pronto serial cable is required to connect the sensor connector of the Rad-57 to the
PC and TrendCom software is required to download the trend data on to the PC. Refer
to Section 10 - Accessories and contact Masimo to acquire these optional items. Patient
monitoring is not possible while trend memory is being transferred to a PC.
A trend data download is initiated using the TrendCom utility which downloads the trend
data and saves it to a space-delimited ASCII text (.out) file.
*See Model Summary for applicable device.
TRENDCOM UTILITY INSTALLATION
Copy the TrendCom utility from the CD onto a PC running MS-Windows.
TRENDCOM UTILITY OPERATION
1. Disconnect patient cable from the Rad-57.
2. Connect the mini-D end of the Rad-57 PRONTO serial cable to the Rad-57 patient
cable connector (See Section 2, Rad-57 front panel controls) and connect the DB-9
end to a COM port on the PC.
3. Turn the Rad-57 on
4. Start the TrendCom Utility
5. Select the appropriate COM port number, if necessary.
6. Push the RETRIEVE TREND button on the TrendCom utility.
Select the desired location and assign a filename for the trend file. Press Save.
7. The Rad-57 will display “dat out” while trend data is being transferred.
A progress bar will advance to indicate the status of the download.
Larger trend files will take longer to download. Transfer time is approximately 20
seconds per hour of trend data.
NOTE: During download of trend information, all normal Rad-57 functions are
unavailable and the keypad is locked, except for the power button.
8. When trend data transfer is complete, close TrendCom and disconnect the Rad-57
from the Rad-57 PRONTO serial cable.
9. Turn the Rad-57 off to exit the trend download mode.
NOTE: USB to serial port adapters are not supported for trend transfer.
NOTE: Enabling trend (setting Trend to “ON”) will erase all trend information in the
Rad-57.