Masimo Corporation 53 www.masimo.com
Chapter 5: Performance Verification Procedures
This chapter covers how to test the operation of Rad-97. No internal adjustment or recalibration is required. These procedures should be done by authorized service
personnel only. Safety checks should be performed at regular intervals or in accordance with local and governmental regulations.
To test the performance of Rad-97 following repairs or during routine maintenance, follow the procedure outlined in this section. If Rad-97 fails any of the described tests,
discontinue its use and correct the problem before returning the unit back to the user.
Battery Test
Battery Replacement Verification
After battery replacement, verify the battery is identified by Rad-97.
1. Connect the power cord to the power entry module on the rear of Rad-97.
2. Verify the battery charge indicator on the home button of Rad-97 front panel illuminates.
• If the battery charging indicator does not illuminate, this indicates the battery may not be properly connected.
3. Press the power button and turn on the device.
4. Verify the battery icon on the Main Screen indicates the battery is charging.
• If the battery icon has a Red "X" through it, this indicates the battery may not be properly connected. Turn off the device and unplug the power cord, then
remove the back panel and recheck the battery cable connection.
Battery Charging
Rad-97 includes a Rechargeable Li-Ion battery Rated at 10.8V 5000 mAh 54.0 Wh. Before use, the Rad-97 battery must be charged completely. As a precaution against
memory effects, it is advisable to completely discharge and fully recharge the batteries periodically.
CAUTION: If the Rad-97 battery has not been charged for 1 month or more, then recharge the battery prior to use.
CAUTION: At low battery warning, connect Rad-97 to AC power to prevent interruption of power.
CAUTION: Additional information on the Rad-97 batteries may be found in the Operator's Manual for Rad-97.
Note: When battery run time is significantly reduced, it is advisable to completely discharge and fully recharge the battery.
To charge Rad-97
1. Plug the AC power cord into the power entry module. Make sure it is securely plugged in.
2. Plug the AC power cord into an AC power source.
3. Verify that the battery is charging:
• When Rad-97 is ON and charging, the AC Power Indicator lightning bolt icon will appear on the screen.
• When Rad-97 is OFF and charging, the Home button will illuminate Orange.
4. When the battery is fully charged:
• When Rad-97 is ON and fully charged, the AC Power Indicator will change to a plug icon.
Touch the AC Power Indicator icon to view battery charge details. Refer to the Operator's Manual for Rad-97 for complete information.
Battery Diagnostics
Battery diagnostics screen displays information about the current health of the battery. To view the Rad-97 Battery Diagnostics screen, perform the following:
1. With Rad-97 On, press/select the battery icon at the top of the Rad-97 Main Screen.
2. Next to Battery Diagnostics, select the View button.
3. When prompted for a password, enter 5664 and the enter key. The Rad-97 Battery Diagnostics screen displays the following information.