Radical Signal Extraction Pulse Oximeter Operator’s Manual 17
3
monitor setup
The Radical Pulse Oximeter stores two types of default values: those that Radical
automatically reverts to after a power cycle, and those that can be changed by the user
and will be remembered after a power cycle.
The following table outlines the default values that Radical reverts to after a power cycle.
OPTION DEFAULT SETTING
ALARM SILENCE Alarm silence time is set to 120 seconds
LCD SCREEN ILLUMINATION (AC power) Illumination set to maximum, level 4
LCD SCREEN ILLUMINATION (battery power) Illumination set to off
The following table outlines the type of settings that the user can change and Radical will
remember after a power cycle.
OPTION DEFAULT SETTING
SpO
2
HIGH/LOW ALARM LIMIT Set to pre-power down setting
PULSE RATE HIGH/LOW ALARM LIMIT Set to pre-power down setting
DISPLAY VIEW AND CONTRAST Set to pre-power down setting
AVERAGING TIME Set to pre-power down setting
FASTSAT Set to pre-power down setting
HOME USE Set to pre-power down setting
INTERFACE ALARMS Set to pre-power down setting
SATSHARE NUMBERS Set to pre-power down setting
POWER SAVE Set to pre-power down setting
SENSITIVITY Set to pre-power down setting
DATE AND TIME FORMAT Set to pre-power down setting
LANGUAGE Set to pre-power down setting
ANALOG OUTPUT Set to pre-power down setting
SERIAL OUTPUT PORT MODE Set to pre-power down setting