EasyManua.ls Logo

Masimo Radical - Replacing Batteries and Fuses

Masimo Radical
78 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60 Radical Signal Extraction Pulse Oximeter Operator’s Manual
9
service/maintenance
WARNING: THE DOCKING STATION BATTERY SHOULD BE INSTALLED AND/OR
REMOVED FROM DOCKING STATION BY QUALIFIED PERSONNEL ONLY.
REPLACING THE BATTERIES
Before installing or removing the battery, make sure the AC power cord is removed
and power to the pulse oximeter is turned off.
To replace the Handheld battery, follow these instructions:
Turn the Radical Handheld off and remove the patient cable connection. Detach the
Radical Handheld from the Docking Station (if docked).
Loosen the closure screw on the battery compartment door and lift off the battery
compartment door cover.
Pull out the battery.
Take a new battery, and place it in the compartment.
Replace the battery door and tighten the closure screw.
Place Handheld into Docking Station, turn on line power and charge battery
according to Section 9, Battery Operation and Maintenance.
CAUTION: FOLLOW LOCAL GOVERNING GUIDELINES FOR PROPER DISPOSAL OF
INTERNAL BATTERIES. DO NOT INCINERATE.

Table of Contents

Related product manuals