48 Radical Signal Extraction Pulse Oximeter Operator’s Manual
7
specifications
radical specifications
PERFORMANCE
measurement range
SpO
2
: 1-100%
Pulse Rate: 25-240 beats per minute (bpm)
Perfusion: 0.02% - 20%
accuracy
Saturation 70% to 100%
No Motion
1
Adults, Pediatrics
±
2 digits
Neonate
±
3 digits
Motion
Adults
2
, Pediatrics
2
±
3 digits
Neonate
3
±
3 digits
Low Perfusion
4
Adults, Pediatrics
±
2 digits
Neonate
±
3 digits
Pulse Rate Accuracy
pulse rate: 25-240 bpm
No Motion
1
Adults, Pediatrics, Neonate
±
3 digits
Motion
2,3
Adults, Pediatrics, Neonate
±
5 digits
Low Perfusion
4
Adults, Pediatrics, Neonate
±
3 digits
resolution
Saturation (%SpO
2
) 1%
Pulse Rate (bpm) 1 bpm
ELECTRICAL
standalone
AC Power requirements: 100-240 VAC, 47-63 Hz
Power consumption: 55 VA
Fuses: 1 Amp, Fast Acting, Metric, (5x20mm), 250V
batteries
Handheld:
Type: NiMH
Capacity: 4 hours
5
Charging time: 3 hours
Docking Station (RDS-1B, RDS-3B):
Type: NiMH
Capacity: 12 hours
5
Charging time: 6 hours