Radical Signal Extraction Pulse Oximeter Operator’s Manual 31
4
operation
This menu allows the user to set the alarm silence period. An
alarm is silenced by pressing the Alarm Silence button on the
front panel.
30, 60, 90,120 SECONDS
The alarm silence duration is set. As an indicator that the alarm
system is silenced, the Alarm Status Indicator is shown as a bell
with a slash through it. A timer is shown next to the bell indicating
the remaining alarm silence duration.
NOTE: The alarm silence period is reset to 120 seconds upon power cycle, except
for when the Radical is set to operate in the Home mode.
ALL MUTE
All patient alarm conditions are silenced. Only system alarms will
be indicated by an audible alarm. As an indicator that the system
is set to All Mute, the Alarm Status Indicator is shown as a
flashing bell with a slash through it.
ALL MUTE WITH AUDIBLE REMINDER
All patient alarm conditions are silenced. Only system alarms will
be indicated by an audible alarm. As a reminder, a single audible
alarm will occur every three minutes. As an indicator that the
system is set to All Mute the Alarm Status Indicator is shown as
a flashing bell with a slash through it.
This menu allows the user to set the alarm volume. Four levels
are available: level 1 being the softest and level 4 being the
loudest. The device retains the Alarm Volume setting upon a
power cycle.
NOTE: For home use, set the alarm level to level 4.
WARNING: IF AN ALARM CONDITION OCCURS WHILE THE ALARM SILENCE
PERIOD IS SET TO ALL MUTE, THE ONLY ALARM INDICATIONS WILL BE
VISUAL DISPLAYS AND SYMBOLS RELATED TO THE ALARM CONDITION. NO
ALARM TONE WILL SOUND.
SILENCE
LEVEL