SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES (CONTINUED)
■
The pulse oximeter can be used during defibrillation, but the readings may be inaccurate for a
short time.
■
This equipment has been tested and found to comply with the limits for medical devices to the
IEC 601-1-2:1994, Medical Device Directive 93/42/EEC. These limits are designed to provide
reasonable protection against harmful interference in a typical medical installation. This
equipment generates, uses and can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful interference to other devices in the
vicinity. However, there is no guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to other devices, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the interference by
one or more of the following measures:
❚
Reorient or relocate the receiving device.
❚
Increase the separation between the equipment.
❚
Connect the equipment into an outlet on a circuit different from that to which the other
device(s) are connected.
❚
Consult the manufacturer for help.
iii Radical Signal Extraction Pulse Oximeter Operator’s Manual