41
Tubing Cartridge Maintenance and Recalibration
®
A clockwise turn will DECREASE the volume
dispensed.
A counter-clockwise turn will INCREASE the volume
dispensed.
Each FULL Revolution of the screw will adjust the
volume level by approximately 8 mg for the ten shots
dispensed (or 0.8 mg/dispense).
6. Repeat the previous step until three consecutive
weights are within the gravimetric range stated
above. Do this with all eight channels. Record
results.
7. Fill the supply container with the appropriate medium
for your plate reader.
8. Make sure that the plate carriage is properly in
position on the unit. Insert a plate in the carriage
and position the nozzle tips to correctly dispense
into the plate.