42
Tubing Cartridge Maintenance and Recalibration
9. Select a program to ll the whole plate to the
appropriate volume and press the START key to
begin the dispense.
10. After the dispense operation has nished, insert the
lled plate into the plate reader for evaluation.
11. After the plate has been read, evaluate the
results. If any CVs for ANY row are greater than
the speci cations for the tube set, the tube set will
need to be adjusted for that channel. Adjust the
calibration by turning the adjustment screws as
described earlier.
12. After all calibration is completed, rinse the tubing
by putting the tube ends in a container holding
distilled water and press and hold the PRIME key
until enough liquid has passed through the tubing
to clear it.
13. Repeat the rinse cycle using alcohol to remove all
water from the line.