EasyManua.ls Logo

Matrix WellMate - Page 61

Matrix WellMate
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
56
Appendices
Syntax DATA Function Comment
O COM OPEN Open is not allowed during
a plate- lling operation;
instrument must be idle.
P±xxxxx xxxxx Pump action:
+ = dispense;
- = pump motor
runs in reverse
Programs volume to
dispense into plate (+) or
returned to supply container
(-).
NOTE: 0.025xxxxx = value
inµl; e.g., P+04000 = 100μL
to be dispensed
P0 Pump: Move
to the original
position (4
different
positions) in the
dispensing way.
This action is to be done
prior to every plate- lling
sequence. It ensures pump
head is at Start position.
P1 Pump action:
Start dispensing
Continue pump action until
P3 command is executed.
P2 Pump action:
Start pumping
in reverse
direction (to
supply)
Reverses pump motor
action. Used for extracting
uid left in tubing back into
supply container.
P3 Stop pump
action.
Stops the pump motor.
Only effective for P1 & P2
commands.
PAL pp Loads program
00 or 10 from
pp depending
on pp See PAW
PAR pp Read a volume
& pattern from
memory
pp - program to write to (00
19) return format: vvvvvx...
x, x count
depends on pp.
See PAW