3. Installation and Commissioning of the SPS Working Station
Page 3–2 Operating Manual MB-SPS-System
©
GmbH – Version 4.3 - 06/2010
3.1 Installation and Commissioning of the Solvent Purification System
The accessories described in this chapter (gases, pressure-reducing valves) are required for
connecting the system. They are not included in the system’s delivery package.
CAUTION
Risk of accident while handling gases!
► Competent and experienced personnel should only carry out
connection of systems.
See also 2.3 Intended Use and 2.4 Improper Use.
3.1.1 Preparation
The customer has to provide the conditions of site locations and the media connections required by
the system.
WARNING
Risk of death, injury and damage, if your exhaust system fails!
It is the responsibility of the customer/ enduser
► to provide a functioning exhaust system and
► To install a detection system, which identifies failure of the
exhaust system and gives appropriate warning.
In case of failure of the exhaust system
► interrupt the operation of the SPS System immediately and
► safeguard the system.
3.1.2 Transport
SPS Compact
The weight of the SPS Compact (without Filling) is 38 kg.
The SPS Compact can be transported by 2 Persons in gripping, bracing and lifting the system's
corpus. Use transport equipment, if necessary.
SPS 800 Manual
The SPS 800 Manual can be moved by rolling. The mounting feet must be screwed in.
3.1.3 Site Location
Selecting the site for an MBRAUN system or any part of a multi-unit system should be carried out
applying the following guidelines. If in doubt, contact the MBRAUN service department for
assistance.