6. Operating the Working Station (SPS 800 Manual and SPS Compact)
Operating Manual MB-SPS 800 Manual and SPS Compact
©
GmbH – Version 4.3 – 06/2010
Page 6–5
6.3.1.2 Pressurizing the solvent reservoir
► Start pressurizing the solvent reservoir (see chapter 5.2.3.2).
The solvent will be pressed to the purification filter columns, passes the getter material and
oxygen and moisture will be filtered out of the solvent.
Dependent on the viscosity of the solvent, the pressurized pressure and the densitiy of the
getter material the passage will take according time.
6.3.2 Removing the solvent at the filling station
NOTE!
Before starting the dispensing procedure, it is mandatory the complete
system is degassed.
The filtered solvent can be removed from
- external at the filling station at the working station or
- within the filling station in a Glovebox (optional)
See following chapter.