5. System Description and Operating: Solvent Reservoir and Sampling Flask
Page 5–16 Operating Manual MB-SPS-Systems
©
GmbH – Version 4.3 – 06/2010
1
(V 3)
(V 4)
At the working Station
► Close the upper Valve (V 3)
► Position the lower Valve (V 4) to Vacuum / N2
2
► Position the upper valve (V 3) slowly to Vacuum.
► Evacuate the flask for at least 1 minute (the required is
dependent on the volume of the flask).
► Check the end pressure at the pressure reading for
vacuum.
3
(V 3)
(V 4)
After required evacuating time,
► position the upper valve ( V 3) slowly to N2 .
→ The flask will be flooded with nitrogen.
► Let the valve in the position N2 for ca. 2 sec.
(the required time is also dependend on
the volume of the flask).
► Evacuating and flooding (step 1 - 4) has to be carried out
at least 3-fold. *)
4
(V 4)
► After the third flooding close valve (V 4)
► After this, evacuating and flooding of the flask is finished.
*) This step is not necessary during initial commissioning.
!
NOTE!
SPS 800 Automatic :
Evacuating the sampling flask is done by an automatic cycle.