24
COMPACT PIEZO LED
5 METHODSANDPRECAUTIONSFOR
DISPOSAL
WARNING: Hospital waste. Treat the
following objects as hospital waste:
• Inserts, when worn or broken;
• Inserts tightening wrench, when worn or
broken.
Disposable materials and materials that carry
biological risk must be disposed of according
to local regulations in force regarding
hospital waste.
WARNING: When handling the inserts,
pay particular attention to the sharp, pointed,
irregular parts to avoid any wounds or
injuries.
compact piezo LED must be disposed of and
treated as waste subject to separate collection.
Failuretocomplywiththepreviouspointscan
result in a penalty pursuant to the directive
on waste electrical and electronic equipment
(WEEE).
The purchaser is entitled to deliver the device
to be disposed of to the dealer who supplies
them with new equipment; at Mectron,
instructions for proper disposal are available.
6 TECHNICALDATA
Devicecompliantwith
Regulation(EU)2017/745
Class lla
ClassicationundertheIEC/EN60601-1
Theclass denitionisdeterminedbythe
manufacturer of the dental unit which
incorporates compact piezo LED.
Partsapplied:typeB(insert)
IP20(appliance)
IP according to indications provided by
the manufacturer of the dental unit
Essentialperformance
AccordingtothestandardIEC80601-2-60
the device has no essential performance
Applianceforintermittentoperation
60sec.ON-30sec.OFFwithirrigation
30sec.ON-120sec.OFFwithoutirrigation
PowerSupply
Power adapter compliant with IEC/EN
60601-1:
• 24V~50/60Hzor
• 32 V
Max.PowerConsumption 40 VA
Fuse T2AL,250V(notincluded)
WorkingFrequency
Automatic Scan
From24KHzto36KHz
Power
Adjustable depending on the type of
installation.