6.1 ElectromagneticCompatibilityIEC/EN60601-1-2
WARNING: Contraindications.
Interferencewithotherequipment
Even if complying with the standard IEC
60601-1-2,compactpiezoLEDmayinterfere
with other devices in the vicinity.
WARNING: Portable and mobile radio
communicationappliancesmayaectproper
device operation.
WARNING: Contraindications.
Interferencefromotherequipment. An
electrical scalpel or other electro-surgical
units near the compact piezo LED device may
interfere with its correct operation.
WARNING:Thedevicerequiresspecic
EMC precautions and must be installed
and commissioned according to the EMC
information provided in this chapter.
WARNING: The technician or the
manufacturer of the dental chair on which
the compact piezo LED has been installed
is responsible for the compliance of the
productandoftheentiresystemtotheIEC/
EN60601-1inlaw.
6.1.1 Guidance and Manufacturer’s Declaration - Electromagnetic
Emissions
compactpiezoLEDisintendedforuseintheelectromagneticenvironmentspeciedbelow.The
customer or user of compact piezo LED should ensure that it is used in such an environment.
EmissionsTest Compliance Electromagneticenvironment–Guidance
RFemissions
CISPR 11
Group1
compact piezo LEDonlyusesRFenergyfor
internalfunction.Therefore,itsFRemissions
are very low and are not likely to cause
any interference with nearby electronic
equipment.
RFemissions
CISPR 11
Class B
compact piezo LED is suitable for use in
all establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltageuctuations/
ickeremissions
IEC61000-3-3
Complies