TECHNICAL DATA
a) The test may be performed at any one power input
voltage within the compact piezo LED RATED voltage
range. If the compact piezo LED is tested at one
power input voltage, it is not necessary to re-test at
additional voltages.
b) All compact piezo LED cables are attached during
the test.
c) Calibration for current injection clamps shall be
performedina150Ωsystem.
d) If the frequency stepping skips over an ISM or
amateur band, as applicable, an additional test
frequency shall be used in the ISM or amateur radio
band. This applies to each ISM and amateur radio
bandwithinthespeciedfrequencyrange.
e) Testing may be performed at other modulation
frequenciesidentiedbytheRISKMANAGEMENT
PROCESS.
f) MEEQUIPMENTandMESYSTEMSwithad.c.power
input intended for use with a.c.-to-d.c. converters
shall be tested using a converter that meets the
specicationsoftheMANUFACTURERoftheME
EQUIPMENTorMESYSTEMS.TheIMMUNITYTEST
LEVELS are applied to the a.c. power input of the
converter.
g) Applicable only to equipment and systems connected
to single-phase a.c. mains.
h) E.g.10/12means10periodsat50Hzor12periods
at60Hz.
i) MEEQUIPMENTandMESYSTEMSwithRATEDinput
currentgreaterthan16A/phaseshallbeinterrupted
oncefor250/300cyclesatanyangleandatallphases
atthesametime(ifapplicable).compactpiezoLED
with battery backup shall resume line power operation
afterthetest.ForMEEQUIPMENTandMESYSTEMS
withRATEDinputcurrentnotexceeding16A,all
phases shall be interrupted simultaneously.
j) MEEQUIPMENTandMESYSTEMSthatdoesnothave
a surge protection device in the primary power circuit
maybetestedonlyat±2kVline(s)toearthand±1
kVline(s)toline(s).
k) Not applicable to CLASS II equipment and systems.
l) Direct coupling shall be used.
m) R.M.S. , before modulation is applied.
n) TheISM(IndustrialScienticandMedical)bands
between0,15MHzand80MHzare6,765MHzto
6,795MHz;13,553MHzto13,567MHz;26,957MHz
to27,283MHz;and40,66MHzto40,70MHz.The
amateur radio bands between 0, 15 MHz and 80 MHz
are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz
to5,4MHz,7MHzto7,3MHz,10,1MHzto10,15
MHz,14MHzto14,2MHz,18,07MHzto18,17MHz,
21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0
MHzto29,7MHzand50,0MHzto54,0MHz.
o) ApplicabletoMEEQUIPMENTandMESYSTEMSwith
RATEDinputcurrentlessthanorequalto16A/phase
andMEEQUIPMENTandMESYSTEMSwithRATED
inputcurrentgreaterthan16A/phase.
p) ApplicabletoMEEQUIPMENTandMESYSTEMS
withRATEDinputcurrentlessthanorequalto16A
/phase.
q) At some phase angles, applying this test to ME
EQUIPMENTwithtransformermainspowerinput
might cause an overcurrent protection device to open.
Thiscanoccurduetomagneticuxsaturationofthe
transformer core after the voltage dip. If this occurs,
theMEEQUIPMENTshallprovideBASICSAFETY
during and after the test.
r) ForMEEQUIPMENTandMESYSTEMSthathave
multiple voltage settings or auto ranging voltage
capability, the test shall be performed at the minimum
andmaximumRATEDinputvoltage.MEEQUIPMENT
andMESYSTEMSwithaRATEDinputvoltagerange
oflessthan25%ofthehighestRATEDinputvoltage
shall be tested at one RATED input voltage within the
range.