PCF Pro User Manual
DOC01851 Page 11 of 74
Part 1, Test Type Acceptable Value Result Comments
Test Through beam sensor basket obstruction along
machine front edge during downward motion halts lifting
system then raises up by approximately 30mm
Test Through beam sensor basket obstruction along
machine front edge during upward motion halts lifting
system
Compressed Air
5.5 – 7 bar
(80 - 100 psi)
Hot Water Hardness (As
specified by WD Manufacturer)
125 ppm (max)
Cold Water Hardness (As
specified by WD Manufacturer)
125 ppm (max)
Hot Water Temperature
50°C (122°F) min
80°C (176°F) max
Cold Water Temperature
5°C (41°F) min
30°C (86°F) max
Hot Water Pressure
2 - 5 bar
(30 – 70 psi)
Cold Water Pressure
2 – 5 bar
(30 - 70 psi)
Part 2, Final rinse water Acceptable Value Result Comments
Rinse water supply Type
RO, DI, Softened
If softened, Hardness
< 125 ppm
Water Pressure
2 – 5 bar
(30 – 70 psi)
Temperature
5°C (41°F) min
80°C (176°F) max
Electrical conductivity
<30µS/cm (RO)
<10µS/cm (DI)
Signed Name Date
Comments /
remedial work