PCF Pro User Manual
DOC01851 Page 38 of 74
Push up the locking flags
ENSURE the thumb ball latches in the down
position before moving the carrier lift
Luer pipes
The PCF Pro basket comes with luer pipe
connections. Luer pipes are connected to
these connections for cleaning of lumen
instruments.
In general, the luer pipes should be replaced
every 12 months. However, depending on use
of the machine, the luer pipes may require
replacement sooner than 12 months.
The user must inspect the luer pipes before
every use and replace if required.
Note: Replace all luer pipes at the same
time.
Caution
• Do not allow Luer pipes to
protrude from the basket.
• If any instruments became
Medisafe connection tubing
during processing, they
should be reprocessed
ensuring that the Luer
connection is firmly seated.
•
basket
hoses, ensure they are not
kinked or ruptured.
The luer pipe clips into the luer connector.
To remove the luer pipe, press the luer
connector clip.