PCF Pro User Manual
DOC01851 Page 51 of 74
5. Check the 4 lid docking seals for any
damage.
If damage is found, please
contact a service engineer (see
section 1-17).
6. Check the six basket docking seals
for any damage or debris.
If damage is found, please
contact a service engineer (see
section 1-17).
7. Check the six basket manifolds for
blockage.
Weekly checks
1. Carry out the daily checks.
2. Check water hardness of hot and
cold water supplies.
3. Check water conductivity of final
rinse water supply.
4. Check condition of mains water
supply hoses.
5. Visually inspect water and drainage
system for leaks.
6. Cleaning efficacy test by residual soil
detection
7. Run a full decontamination
procedure (see section 3-14).
8. Test operation of safety devices (see
section 3-7-10).
Note: The use of a “Process Challenge Device
for Minimally Invasive Surgery” may be
required according to your local authority
guidelines, and the cleaning indicator
manufacturer’s instructions.
Ref:
EN ISO15883-1:2009