EasyManua.ls Logo

Medisafe PCF Pro - Starting a Cycle

Medisafe PCF Pro
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PCF Pro User Manual
DOC01851 Page 42 of 74
Starting a Cycle
The load sequence is optimised for hands free
operation using the Load foot pedal to step
through an automated sequence:
1. Confirm program selection (to
change see section 3-5)
Pinch points are created during lid
articulation. Keep hands clear
when opening or closing the lid.
2. Press and hold the Load foot pedal
until the carrier lift is fully open
If there is insufficient chemical to
complete the cycle a warning will be
shown:
Replace chemical(s) (see section 3-
6-2) and press OK to continue, or
press Cancel to abort the cycle start
3. Insert the bottom basket into machine
and latch (see section 3-7-2)
4. Press and hold the Load foot pedal
until the carrier lift indexes down to the
next level and stops
5. Insert the middle basket into machine
and latch (see section 3-7-2)
6. Press and hold the Load foot pedal
until the carrier lift indexes down to the
next level and stops
7. Insert the top basket into machine and
latch (see section 3-7-2)
8. Press and hold the Load foot pedal
until the carrier lift is closed, and the
cycle will start
Note: the touch screen controls may be used
to bypass the automated sequence (see
section 3-7-6)

Table of Contents