Other Faults
Problem Possible Causes and solutions
Poor Load Cleaning
Insufficient detergent injected. Contact Medisafe
Check that an approved chemical is being used
Instrument basket is overloaded. Adjust load
Instruments are excessively soiled
Instruments have been left soiled for more than two hours
Verify facility power meets Unit specifications
Water Leaking from Lid
Area
Check instruments are lying flat in the basket and not pointing
up at the lid
Check lid seal is undamaged
Excessive Fill Time
Check incoming water pressure is above 2 bar (29 psi)
Check water inlet valve is not blocked (beware of hot water
temperature!)
Low Irrigation Port Flow
Rate
Check all connectors are fully inserted
Clean tank filter
Check basket manifold ports are not blocked
Check tank docked ports are not blocked
Verify facility power meets Unit specifications
No Power to Unit or
Display is off
Installation Main disconnect switch is in the OFF position
Turn main disconnect switch to the ON position
Facility power is incorrect for unit. Correct facility power to
meet unit requirements
Internal Circuit breaker has tripped, contact Medisafe
Access to the load in the
event of a power failure or
stuck lid mechanism
The PCF Pro contains a manual override to allow the lid to be
opened, this must be carried out by trained service personnel.
Please contact Medisafe.