Re-Wash Cycle
The PCF pro can be configured with a re-wash
program - a water and time saving feature.
This allows the wash water and chemical to be
reused. At the end of the wash, the user is
allowed to open the lid, add new instruments,
and repeat the wash stage using the same
water up to ten times.
Note: if the water temperature exceeds the set
point by more than 10°C the cycle will be
aborted (see section 3-9-10).
At the end of a wash stage press Yes on the
screen to reuse the water again.
Unload the instruments (see section 3-9-9)
then load more instruments for cleaning (see
section 3-8).
Temperature of the water within
the unit can reach 60°C
(140°F), when opening the lid
and reloading instruments take
care and wear appropriate PPE.
Press Reload Complete on the screen to
begin another wash stage
To end the re-wash process, once the wash
stage is complete press No on the screen to
drain the water and end the cycle.