PCF Pro User Manual
DOC01851 Page 50 of 74
Caution
Check inside the spray arm and clear
any debris.
To ensure correct spray direction the
spray arms are handed so they
cannot be inserted incorrectly:
To replace the spray arm, line up the
slot to the internal hub pin, push in
and then rotate clockwise:
6. Check the condition of basket
irrigation port Luer hoses, ensure
they are not kinked or ruptured.
In accordance to ISO15883, the
international standard for the performance
of Washer-Disinfectors, we advise the
checks stipulated below. However, each
hospital may follow its own checks to
comply with the applicable standards.
Daily Checks
1. Check that there is sufficient
chemical for the days planned use
2. Check and clean all internal surfaces
above the water line.
3. Check and clean all external
surfaces including lid seal.
4. Check lid and lid seal for any
damage.
If damage is found, please
contact a service engineer (see
section 1-17).