PCF Pro User Manual
DOC01851 Page 39 of 74
The luer pipe disconnects from the luer
connector as below.
Remove the luer pipe by lifting it away from the
luer connector.
Instruments
Each basket is equipped with twenty ports for
maximum pressure irrigation. These are
divided equally between each side to prevent
crossovers or shadowing of instruments.
When selecting Instruments for the PCF Pro:
Always confirm compatibility with the
instrument manufacturer.
Always follow the manufacturer’s
instructions if processing instruments
with special characteristics e.g. fibre
optics or rigid optics.
Whether pre or post processing
instruments, Always follow the
manufacturer’s instructions.
Instrument Loading Precautions
When loading instruments into a basket
observe the following precautions:
Do not place items directly into the
wash tank or allow them to make
contact with the tank sides.
Do not load basket with instruments
pointing at the lid.
Do not overload the basket, maximum
load of 10kg (22lbs) per basket.
Do not allow Luer pipes to protrude
from the basket.
Always use a Full DIN Tray Standard
to hold instruments (M20044).
Always attach tubing to a suitable size
distal connector for instruments being
cleaned or to a quick connect port if
the instrument has one.
Always open box jointed instruments.
Always ensure taps are open on
cannulated instruments.
Always dissemble instruments that
are designed to be taken apart.
Luer pipe disconnects
Lift luer pipe to remove