PCF Pro User Manual
DOC01851 Page 57 of 74
Error
Code
Error Message Possible Causes and Solutions
14
Chamber Start Temperature
too High
Check the Cold water temperature is not above 35°C
The chamber has not cooled sufficiently since the last wash.
Leave more time between washes for the machine to cool down.
15
Chamber Temperature Out of
Range
Switch the machine off then on again
18 19
51
Level Sensor Invalid Repeat cycle
20
User Aborted Cycle Repeat cycle
21
Low Main Pump Pressure
Check tank filter is clear/clean
Check baskets are docked properly
Check basket ports are not blocked
Check basket docking seals clean
22
High Main Pump Pressure
Check baskets are docked properly
Check basket ports are not blocked
Check basket docking seals clean
23
Temperature Set Point
Exceeded
Repeat cycle
38
Compressed Air Absent or
Pressure Low
Check compressed air to the machine is turned on
41 43
45 46
Sonic Generator Fault Switch the machine off then on again
42
Maximum Cycle Time Expired
Check water supply flow rate is sufficient
Check tank filter clear, ensure drain is not blocked
44
Sonic Cleaning Water Level
Low
Check for visual leaks
48
Emergency Stop Pressed
Determine reason why emergency stop was pressed before pulling the
button to release
49 57
58
Microcontroller Error Switch the machine off then on again
50
Internal Ambient Temperature
too High
Switch the machine off then check for blockages in the ventilation
holes
Switch the machine on again after an hour
52
Real Time Clock Read Error Switch the machine off then on again
53
24V Power Supply Voltage
Out of Range
Switch the machine off then on again
54
Overfill occurred during cool
down process
Check tank filter clear, ensure drain is not blocked
Repeat cycle
56
Low Wash Irrigation Pressure
Check tank filter is clear/clean
Check baskets are docked properly
Check basket ports are not blocked
Check basket docking seals clean
59
Unable to Store Cycle Log Repeat cycle