PCF Pro User Manual
DOC01851 Page 24 of 74
Water Supply
Water Hardness
(CaCO
3
), Quality & Purity
<125ppm with water quality and purity tested and approved for Disinfection as per
EN ISO15883-1:2009 (item 6.4.1 and 6.4.2)
Pressure
2 – 5 bar (
30 – 70 psi
) for all water supplies
Flow Rate 1230 l/m (3 8 US gal/m)
Temperature:
Cold Water 5 – 30°C (41 86°F)
Hot Water 50 – 8C (122140°F)
RO Water 5 – 80°C (
41 140°F
)
Connection
1x shut valve with ¾in BSP male thread for hose tail (
Hot water
)
1x shut valve with ¾in BSP male thread for hose tail (Cold water)
1x shut valve with ¾in BSP male thread for hose tail (RO water)
Position
Within 1m (40in) of proposed installation
Drainage System
Type
Gravity
Drain Connection
54mm (2.125in) OD outlet from machine
Drain must be fitted with suitable traps
Position
Drain entry and the entire drain hose must be below the machine drain outlet
Drain gulley pipe ID at least 100mm (4in) to conform with EN ISO15883
Capacity > 180ltr/m (47.5 US gal/m)
Overflow
Internally connected within the machine to the drain pipe, downstream of the drain
valve
Effluent Temp
95°C (
203°F
)
Compressed Air Supply
Pressure
5.5 – 7 bar (80 100 psi)
Quality
If medical grade compressed air is not used, an Eliminizer needs to be included at
time of order
Connection 8mm (0.315in) dia. Quick fit hose connector
Printer
Dimensions W:230mm (9in) x D:280mm (11in)
Position
Above or to the side and at least within 1m (40in) from the machine, where it will not
be affected by spillage, leakage or condensation from the equipment.
Supply Connection 90 - 230Vac 50/60Hz (single phase).
Position of printer
power supply
Within 1m (
40in
) from the machine, where it will not be affected by spillage, leakage
or condensation from the equipment.

Table of Contents