81 Medit T-series User Guide
Make sure to unplug the power cable before examining the equipment.
Do not spray the scanning spray into the scanner.
Do not touch the mirror inside the scanner.
Any parts replacement should only be done by service personnel.
Disposal
All components are designed to uphold with following Directives:
RoHS, Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic
Equipment (2011/65/EU).
WEEE, Waste Electrical and Electronic Equipment Directive. (2012/19/EU).
Electrical Safety
The MD-ID0400 system is a Class I device.
To prevent electric shock, the scanner must only be connected to a power source with a protective
earth connection. If you are unable to connect the supplied plug of the scanner into the main outlet,
contact a qualified electrician to replace the plug or outlet. Do not try to circumvent these safety
guidelines.
There is a risk of electric shock if you attempt to access the internal parts of the scanner. This only
should be done by the service personnel.
Do not connect the intraoral scanner to a regular power strip or extension cord as these connections
are not as safe as grounded outlets. Failure to adhere to these safety guidelines may result in the
following hazards:
The total short circuit current of all connected equipment may exceed the limit
specified in EN / IEC 60601-1.
The impedance of the ground connection may exceed the limit specified in EN / IEC
60601-1.
You should only disconnect the scanner from the power supply via its power cord.
Before disconnecting the power cord, make sure to turn off the power on the device using the
power switch on the main unit.
Only use the power adaptor supplied by Medit. Using other power adaptors may result in damage
to the system.
Avoid pulling on the communication cables, power cables, etc. used in the scanner.
The radiation characteristics of the MD-ID0400 makes it suitable for use in industry and Hospitals
(CISPR 11 class A). If the MD-ID0400 system is used in a residential environment (CISPR 11 class B), it
may not provide adequate protection from radio frequency communications.