www.megger.comPAT400 Series
47
If a Bond or Continuity test fails then the user is always allowed to repeat the test. If the test is repeated then the failed results is
disregarded. You may wish to repeat a test in this way if the test probe did not make proper contact with the appliance.
Insulation and Flash test can also be repeated by selecting the ∞ symbol in the Cycles column. This feature is convenient for testing
appliances that have multiple exposed conductive areas.