EasyManua.ls Logo

MicroPace StimLab - Stimlab Configuration; Description of System; How Supplied; Figure 6 Stimlab System Components

Default Icon
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
USER INSTRUCTION MANUAL
Page 20 of 49 ENGLISH
8
8
STIMLAB CONFIGURATION
8.1 Description of system
The Micropace StimLab™ EP Stimulator System is a computerized Cardiac EP diagnostic stimulator
system based on the EPS320 Cardiac Stimulator with a new hardware platform to support a remote
bedside slave monitor and controller located up to 17m from the central installation. The remote
bedside controller displays all Stimulator settings and allows their supervision and adjustment if
necessary by the scrubbed physician.
8.2 How Supplied
Main Components of the StimLab™ system comprise of the EPS320 Stimulus Generator Unit (SGU),
the Computer Cabinet, the Local and the Remote Bedside Controllers.
Figure 6 StimLab™ System Components
Local Station
Touch Screen,
MP3114
Remote Bedside
Controller,
MP3168
(mounted on a roll
stand or trolley,
MP3183)
Bedside Controller
hot pluggable
connector
Four Output
Connection,
Box, MP3086
EPS320 Stimulus
Generator Unit
(SGU), MP3008
Computer
Cabinet ,
MP3167

Table of Contents